| For biocompatible modules use bio / biocompatible parts only! Do not mix with bio-inert parts. |
When
|
Tools required
Qty. | p/n | Description | |
|---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]()
|
| Hex key 4 mm (part of 5023-2524 (Hex Key Set) ) |
1 | ![]()
|
| Cleaning tissue and appropriate solvent like isopropanol or methanol |
Parts required
Qty. | p/n | Description | |
|---|---|---|---|
1 | ![]() | 5067-5918 | Flush Head Seal 500 µL (for G7167C, and G4767A) OR |
1 | ![]() | G5668-60494 | Seal 500 µL Bio-inert (for G7137B) |
Prerequisites
|
Place the new seal to the tip of the piston.
Close the front door.
In the Local Controller close Change Metering Device.
OR
In the Agilent Lab Advisor software Change Metering Device click Back.
For G7167C and G7137B, proceed with the subsequent steps.
Perform Hydraulic Path Leak Test (before FW D07.44 and LabAdvisor 2.23).
OR
Perform Flush Head Function Test (from FW D07.44 and LabAdvisor 2.23).