Error ID: 25034
The autosampler failed to complete the alignment correctly.
Probable cause | Suggested actions | |
|---|---|---|
1 | Mechanical obstruction of the sample handler Mechanical obstruction of the sample handler. |
Ensure unobstructed movement. |
2 | Defective sample handler motors Defective sample handler motors. | Please contact your Agilent service representative. |
base-id: 3803574027
id: 36028800822537995