Error ID:
Quick Noise Test Evaluation
Probable cause | Suggested actions | |
|---|---|---|
1 | Insufficient lamp warm-up time Insufficient lamp warm-up time. |
Allow detector and UV-lamp turned on for at least 2 hours. |
2 | Absorbing solvent or air bubble in flow cell Absorbing solvent or air bubble in flow cell. |
Ensure the flow cell is filled with water, and free from air bubbles. |
3 | Dirty or contaminated flow cell Dirty or contaminated flow cell. |
Flush flow cell
Clean the flow cell as described in Clean the Max-Light Cartridge Cell. Clean the flow cell as described in Maintenance of Standard, Semi-Micro or Micro Flow Cell. |
4 | Old UV-lamp Old UV-lamp. |
Exchange the UV-lamp. |