When
|
Parts required
Qty. | p/n | Description | |
|---|---|---|---|
1 | ![]() | G7123-60200 | Hg-Xenon Lamp |
Prerequisites
|
WARNING
Risk of fire and injury from hot surface.
If the detector has been in use, the lamp may be hot.
WARNING
Injury from glass cuts or mercury poisoning.
The lamp is under high pressure and might burst if not handled with care. The lamp contains mercury (Hg) which might spill if the lamp breaks.
WARNING
Injury from glass cuts or mercury poisoning.
Glass from a broken lamp might cause injury and liquid mercury can cause poisoning over prolonged exposure.
WARNING
Injury from hot surface, glass cuts or mercury poisoning.
In case a lamp breaks when installed in the detector, evaporated mercury can cause poisoning. Hot glass shards from a broken lamp might cause burns or injury.
NOTE
The Hg-Xenon Lamp (G7123-60200) shall only be used as a light source for the Agilent G7123B Fluorescence Detector.
NOTE
Maximum safe daily UV/blue light energy dose at a distance of 50 cm (according to IEC 62471) is reached after 8 hours if detector is operated with opened lamp service door.
NOTE
The used lamp is different to the Xe flash lamp, which is used in G7121A/B Fluorescence Detectors.
NOTE
The lamp house cover includes a magnet.
Open the doors.

Remove the safety grid by pulling on the metal handle.

Pull the metal rod towards the front, so the lamp housing gets turned by 90 degrees.
Press on the lamp cap and turn it counterclockwise until it comes to a stop.
Take off the lamp cap.
Carefully pull out the Hg-Xe lamp.
Place the old lamp in the original packaging immediately after exchange.
Carefully insert the new lamp and make sure the notch on the lamp ceramics matches the metal pin on the lamp housing. The position of the notch is indicated on the lamp ceramics.
WARNING
Injury from glass cuts or mercury poisoning.
Do not use lamps with a runtime over 3000 hours.
Install the lamp cap. Pay attention to the position of the metal flaps and the notches. Push in, then turn the lamp cap clockwise until it comes to a stop.

Push the metal rod on the lamp housing up-/backwards, so the lamp housing turns by 90 degrees until it comes to a stop and gets held in place by a magnet.

Re-install the safety grid. Insert at the bottom, then the top gets held in place by a magnet.

Close the doors.

Power on the module.
Perform the Replace the Hg-Xe Lamp procedure in Lab Advisor.
Ignite the lamp.
Install 5022-2159 (Restriction capillary, SST 0.12 mm ID, 2 m long) .
Flush the system with fresh, clean water.
Perform a wavelength calibration.
Run a Lamp Intensity Check.