Safety Information Related to Installation
WARNING
Absence of ground connection
The absence of ground connection can lead to electric shock
or short circuit.
- Never operate your
instrumentation from a power outlet that has no ground
connection.
WARNING
Electrical shock hazard
Solvents may damage electrical cables.
- Prevent electrical cables from getting in contact with
solvents.
- Exchange electrical cables after contact
with solvents.
WARNING
Heavy weight
The module is heavy.
- Carry the module at least with 2 people.
- Avoid back strain or injury by following all precautions for lifting heavy objects.
- Ensure that the load is as close to your body as possible.
- Ensure that you can cope with the weight of your load.
WARNING
Inaccessible power plug.
In case of emergency it must be possible to disconnect the
instrument from the power line at any time.
- Make sure the power connector of the instrument can be easily reached and
unplugged.
- Provide sufficient space behind the
power socket of the instrument to unplug the cable.
WARNING
Incorrect line voltage at the module
Shock hazard or damage of your instrument can result if the devices are
connected to line voltage higher than specified.
- Connect your module to the specified line voltage.
WARNING
Hazard of electrical shock or damage of your
instrumentation
can result, if the devices
are connected to a line voltage higher than specified.
- Connect your instrument to the specified line voltage
only.
WARNING
Electrical shock hazard
The module is partially energized when switched off, as long
as the power cord is plugged in.
The cover protects users from
personal injuries, for example electrical shock.
- Do not open the cover.
- Do not operate the
instrument and disconnect the power cable in case the cover has any signs of
damage.
- Contact Agilent for support and request an
instrument repair service.
WARNING
Unintended use of power cords
Using power cords for unintended purposes can lead to
personal injury or damage of electronic equipment.
- Never use a power cord other than the one that Agilent shipped with this
instrument.
- Never use the power cords that
Agilent Technologies supplies with this instrument for any other
equipment.
- Never use cables other than the ones
supplied by Agilent Technologies to ensure proper functionality and compliance
with safety or EMC regulations.
WARNING
Use of unsupplied cables
Using
cables not supplied by Agilent Technologies can lead to damage of the
electronic components or personal injury.
- Never use cables other than the ones supplied by Agilent Technologies to
ensure proper functionality and compliance with safety or EMC
regulations.
CAUTION
Condensation within the module
Condensation can damage the system
electronics.
- Do not store, ship or use your
module under conditions where temperature fluctuations could cause condensation
within the module.
- If your module was shipped in
cold weather, leave it in its box and allow it to warm slowly to room
temperature to avoid
condensation.
CAUTION
"Defective
on arrival" problems
If there are signs of
damage, please do not attempt to install the module. Inspection by Agilent is
required to evaluate if the instrument is in good condition or
damaged.
- Notify your Agilent sales and
service office about the damage.
- An Agilent
service representative will inspect the instrument at your site and initiate
appropriate actions.
CAUTION
Solvent incompatibility
The solvent DMF (dimethylformamide) leads to corrosion of the leak sensor. The material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
- Do not use DMF as mobile phase.
- Check the leak sensor regularly for corrosion.
- Do not place metal parts on the leak panel or the leak sensor.
NOTE
Agilent recommends that you install the HPLC instrument in the InfinityLab Flex Bench rack. This option helps to save bench space as all modules can be placed into one single stack. It also allows to easily relocate the instrument to another lab.