Error ID: 2044
The active-inlet valve in the module has drawn excessive current causing the inlet-valve electronic fuse to open.
Inlet-Valve Fuse 0: Pump channel A
Inlet-Valve Fuse 1: Pump channel B
One of the active-inlet valves in the module has drawn excessive current causing the inlet-valve electronic fuse to open.
Probable cause | Suggested actions | |
|---|---|---|
1 | Defective active inlet valve Defective active inlet valve. |
Restart the module. If the error message appears again, exchange the active inlet valve. |
2 | Defective connection cable (front panel to LPM board). Defective connection cable (front panel to mainboard). | Please contact your Agilent service representative. |
3 | Defective main board Defective mainboard. | Please contact your Agilent service representative. |