Error ID: 25226
The autosampler failed to complete injection sequence correctly. The autosampler can move and draw sample from the draw position and generates the error message.
Probable cause | Suggested actions | |
|---|---|---|
1 | Sample container is not installed correctly in the pallet Sample container is not installed correctly in the pallet. |
Check if the sample container is installed correctly. |
2 | Sample container definition in the CDS is not correctly Sample container definition in the CDS is not correct. |
Check if the correct sample container is selected in the CDS.
Verify if the dimension of the sample container match the database of your CDS. |
3 | Sample handler not aligned correctly Sample handler not aligned correctly. |
Check if the sample handler can move freely. Do an autoreferencing (the needle assembly must be installed in the needle parkstation during this procedure). If this will not help: Please contact your Agilent service representative. |