Error ID: 30068
Parameter 0: Excitation ADC for motor current measurement failed.
Parameter 1: Emission ADC for motor current measurement failed.
Probable cause | Suggested actions | |
|---|---|---|
1 | ADC for current measurement on mainboard defective ADC for current measurement on mainboard defective. | Please contact your Agilent service representative. |