Error ID: 30053
The lamp was shut down due to a critical lamp condition.
Parameter 0: High lamp temperature detected.
Parameter 1: Lamp voltage/current issue detected.
Probable cause | Suggested actions | |
|---|---|---|
1 | Over temperature, lamp was shut down because of too high temperature Over temperature, lamp was shut down because of too high temperature. | Check lab conditions. If problem recurs, contact your Agilent service representative. |
2 | Abnormal voltage/current, lamp was shut down caused by surveillance or lamp failed cause of unknown reason Abnormal voltage/current, lamp was shut down caused by surveillance or lamp failed cause of unknown reason. | Exchange lamp. If problem recurs, contact your Agilent service representative. |
base-id: 13925777803
id: 18014412435259787