Probable cause | Suggested actions | |
|---|---|---|
1 | ProbableCause Air in the primary pump chamber |
Check for air in the solvent inlet lines and the Tuning signal: Purge the lines, condition and if necessary prime the pump head |
2 | ProbableCause Abort due to over pressure |
Check solvent and solvent settings Purge and condition the pump head with water |
3 | ProbableCause Leak in inlet valve |
Check for moving air bubbles in tubing directly to the inlet valve: Purge the lines with water to remove dirt Knock at the valve, clean it or replace it |
4 | ProbableCause Outlet valve not properly assembled |
Retighten the outlet valve Check the position of the gold seal |
5 | ProbableCause Leaky piston seals and/or position-dependent leaks on the piston |
Remove the SW tubes from the support ring and check for leaks
Replace the piston seals and clean the pistons: Ensure that seals are lubricated when pushed in Use abrasive mesh > 5000 grit |