When
|
Tools required
Qty. | p/n | Description | |
|---|---|---|---|
1 | ![]() | 8710-0510 | Open-end wrench 1/4-5/16 inch |
1 | ![]() | 8710-1924 | Open-end wrench 14 mm |
1 | ![]() | 5067-5688 | Torque wrench 1 - 25 Nm with 14 mm wrench |
Parts required
Qty. | p/n | Description | |
|---|---|---|---|
1 | ![]() | G7131-60028 | Outlet Valve biocompatible |
Prerequisites
|
Close the shut off valves to avoid solvent leaks.
Insert the new outlet valve and tighten it using a torque wrench with a 14 mm bit set to 10 Nm.
Reconnect all hydraulic connections, and power up the pump.
Open the shutoff valves and purge the system to remove air.
base-id: 4133775243