Error ID: 2687
The electronic fuse on the recycle valve has been activated.
Probable cause | Suggested actions | |
|---|---|---|
1 | Short in recycle valve circuit Short in recycle valve circuit. | Power cycle the module. |
2 | Defective recycle valve Defective recycle valve. | Replace recycle valve. Please contact your Agilent service representative. |
3 | Defective main board Defective mainboard. | Please contact your Agilent service representative. |