Error ID: 1071
This temperature sensor (located on the detector main board) delivered a value outside the allowed range. The parameter of this event equals the measured temperature in 1/100 centigrade. As a result the temperature control is switched off.
Probable cause | Suggested actions | |
|---|---|---|
1 | Dirty or defective sensor. Defective sensor or main board. | Please contact your Agilent service representative. |
2 | Detector is exposed to illegal ambient conditions Detector is exposed to illegal ambient conditions. |
Verify that the ambient conditions are within the allowed range. |
base-id: 3284377995
id: 18014401793859979