Error ID: 24001
The valve drive was not able to operate the valve head correctly. Either due to mechanical reasons or the movement couldn’t be detected correctly.
Probable cause | Suggested actions | |
|---|---|---|
1 | ProbableCause Mechanical problems. Friction too high or blockages on the valve drive’s motor or on the valve head. |
Check valve head for correct installation.
Try to identify the source of trouble by installing a different valve head if possible. Please contact your Agilent service representative. |
2 | ProbableCause Defect Sensor on the Valve Drive Motor. |
Check valve head for correct installation.
Try to identify the source of trouble by installing a different valve head if possible. Please contact your Agilent service representative. |