Error ID: 25158
Probable cause | Suggested actions | |
|---|---|---|
1 | Capillaries are not tightened properly Capillaries are not tightened properly. | Re-tighten capillary connections. Use Hydraulic Path Leak/Sampler Leak test for detailed troubleshooting. |
2 | There is too much air in the sampler hydraulic path There is too much air in the sampler hydraulic path. | Prime S1-S3 solvent lines properly, make sure that there is no air in the solvent lines before Multiwash block. Use Hydraulic Path Leak test for detailed troubleshooting. |
3 | Solvent is too compressible Solvent is too compressible. | Consider replacing the solvent used for Feed Injection. |
4 | Not enough sample in vial and air has been drawn Not enough sample in vial and air has been drawn. | Select appropriate Vial or Vial inserts (recommended 2 ml vial inserts to be used: 5183-2086 (Deactivated glass) , 5181-3377 (Non-Deactivated glass) ). Revise sample draw settings (Draw/Ejection speed, Offset, Vial/Well bottom sensing). |